Background: Accurate flow diverter (FD) sizing is critical for successful intracranial aneurysm (IA) treatment. We evaluated the predictive performance of the Ankyras FD simulation platform within a standardized simulation based clinical workflow using three-dimensional (3D) DynaCTA centerline analysis as an advanced validation reference standard. Methods: A retrospective analysis of a prospectively maintained institutional database of Ankyras guided IA treatments between April 2024 and October 2025 was performed. 3D rotational angiography data generated Ankyras simulations for treatment planning, device selection, and intraoperative guidance. Simulated length (SL) and proximal landing zone (PLZ) were compared with post-deployment length (DL) from DynaCTA centerline measurements (3D Slicer). Accuracy was assessed using correlation coefficients, mean absolute error (MAE), and Bland-Altman analysis. Results: 54 consecutive patients (48 women, mean age 54.2 years) with 63 IAs underwent 57 pipeline embolization device (PED) treatments. All achieved procedural success without intraoperative complications; two required adjunctive stent placement for malapposition. Mean SL was 21.44±3.53 mm and DL was 21.49±3.52 mm (n=46), with a strong SL-DL correlation (r=0.91), minimal bias (-0.05 mm), and 1.24 mm MAE. PLZ demonstrated a strong correlation (r=0.94), low bias (0.14 mm), and minimal MAE (1.34 mm) between Ankyras and the deployed values (n=45). Bland-Altman analysis showed narrow agreement limits without evidence of proportional error. Conclusions: Using a robust 3D DynaCTA centerline based validation, Ankyras demonstrated high accuracy for PED deployment length and landing zone prediction within a standardized clinical treatment workflow of pre-procedural planning, intraoperative guidance, and post-deployment assessment.
Background: Accurate device sizing is crucial for successful flow diverter (FD) therapy in intracranial aneurysms. This study assesses the accuracy and clinical utility of the Ankyras virtual simulation software (Mentice AB, Gothenburg, Sweden) in predicting FD length across multiple device types and manufacturers. Methods: We retrospectively analyzed 193 FDs (7 device types) deployed in 180 patients with 230 intracranial aneurysms. Simulation-based prediction of effective lengths (simulated length (SL)) and nominal manufacturer specifications (labeled length) were compared with measured in vivo lengths (measured length (ML)). Performance was evaluated using simulation accuracy (SA), absolute error (AE), relative error (RE), length ratio, and correlation analysis. Results: Virtual simulation demonstrated good clinical usability, achieving a mean absolute deviation of only -1.38 mm at the proximal landing zone compared with ML. Simulation-based predictions showed significantly superior accuracy compared with nominal manufacturer specifications (SA 89.6±11.3% vs 81.8±13.7%, P<0.001), with mean AE reduced by 58% (-1.38 mm vs -3.34 mm) and mean RE by 42% (10.5% vs 18.2%). Strong correlations between SL and ML (r=0.900) validated predictive reliability across all tested device types. Centerline correction technology further enhanced parameters such as SA to 94.8±9.5%. Conclusion: Virtual simulation-based prediction of effective FD length using the Ankyras software showed reliable and clinically meaningful results, enabling accurate estimations of the proximal landing zone and overall length. Ankyras, along with comparable simulation platforms, may provide considerable potential to facilitate FD selection and implementation, especially in anatomically complex cases.
Flow diverters (FDs) have revolutionized intracranial aneurysm (IA) treatment. Proper FD sizing is crucial for aneurysm occlusion and complication prevention. Ankyras (Mentice, Gothenburg, Sweden) is a device-specific sizing software. We evaluated Ankyras’ performance in predicting the final deployment of the Pipeline™ Vantage Embolization Device with Shield Technology™ (PVST) (Medtronic Neurovascular, Irvine, California, USA).
Flow diverter porosity directly influences the blood flow reduction at the aneurysm neck level and the anatomical result of the treatment. In this research, we present and compare three methodologies to determine the local porosity of deployed flow diverters.
Purpose: During follow-up of patients treated with WEB devices, shape changes have been observed. The quantitative three-dimensional measurement of the WEB shape modification (WSM) would offer useful information to be studied in association with the anatomical results and try to better understand mechanisms implicated in this modification phenomenon.
Flow diverters are widely extended in clinical practice for intracraneal aneurysms treatment. They are formed by a dense mesh of braided wires that partially occludes the aneurysm neck and restores the blood flow into the parent vessel. The occlusion degree is highly dependant on the distribution of the wires under the aneurysm, which is affected by the vessel geometry. Nowadays, there are no clinical indicators of the covering ratio once the flow diverter is deployed. We propose a novel method for the simulation of the flow diverter local porosity before its deployment into the parent vessel. We validate the method on curved silicon models, obtaining a correlation of 0.9 between the simulated values and the measured porosity on the deployed flow diverter.
Intrasaccular devices, like Woven EndoBridge (WEB), are novel braided devices employed for the treatment of aneurysms with a complex shape and location, mostly terminal aneurysms. Such aneurysms are often challenging or impossible to treat with other endovascular techniques such as coils, stents, flow diverter stents.
Flow diverter (FD) devices show severe shortening during deployment in dependency of the vessel geometry. Valid information regarding the geometry of the targeted vessel is therefore mandatory for correct device selection, and to avoid complications. But the geometry of diseased tortuous intracranial vessels cannot be measured accurately with standard methods. The goal of this study is to prove the accuracy of a novel virtual stenting method in prediction of the behavior of a FD in an individual vessel geometry.
Sizing of flow diverters (FDs) stent in the treatment of intracranial aneurysms is a challenging task due to the change of stent length after implantation
Intra-saccular devices (ID), developed for the treatment of bifurcation aneurysms, offer new alternatives for treating complex terminal and bifurcation aneurysms. In this work, a complete workflow going from medical images to post-treatment CFD analysis is described and used in the assessment of a concrete clinical problem.
Flow-Diverter (FD) porosity has been pointed as a critical factor in the occlusion of cerebral aneurysms after treatment. Objective: Verification and Validation of computational models in terms of predictive capacity, relating FD porosity and occlusion after cerebral aneurysms treatment.
Intra-saccular devices (ID) are novel braided devices used for complex intracranial aneurysms treatment. Treatment success is associated with correct device size selection. A technique that predicts the ID size within the aneurysm before intervention will provide a powerful computational tool to aid the interventionist during device selection.
Flow diverters are widely extended in clinical practice for intracraneal aneurysms treatment. They are formed by a dense mesh of braided wires that partially occludes the aneurysm neck and restores the blood flow into the parent vessel. The occlusion degree is highly dependant on the distribution of the wires under the aneurysm, which is affected by the vessel geometry. Nowadays, there are no clinical indicators of the covering ratio once the flow diverter is deployed.
We are here to guide and help you find the solution that is the best fit for your unique needs. Fill in the form to connect with us and a suitable solution expert will get back to you as soon as possible
Rambla Catalunya 53, 4-H, 08007, Barcelona, Spain.
+34 933 28 39 64
Ankyras is manufactured by Mentice Spain S.L. and distributed by Mentice subsidiaries and partners in approved markets. Mentice Spain S.L. is ISO 13485 certified with number MD 784027 Ankyras is CE marked under MDR (MDR 784032) and holds UKCA certification (UKCA 784033). It is 510(k) cleared by the FDA (K230006) and notified to ANVISA (registration number 81118630030).
* This feature may not be available in your country, please check with your local Mentice contact.